Process Stream Sampling Valve

Valco and Global FIA have developed a new multi-position selection valve which has a unique flow-through port that can be used for sampling process streams. The flow-through feature ensures that the sample line is flushed out before sampling, presenting a fresh sample at the sampling port. The remaining ports can be used for normal SIA functions, which makes the valve ideal for on-line process SIA.

The accompanying diagram illustrates the three possible actions that this new valve accommodates.

Position A:
The sample stream flows under process pressure or propelled by a pump through the sample port (2) in a by-pass position sending flushed sample to waste. During this stage, the sample line is refreshed. In this position, the selected port (1) is simultaneously being used for an SIA function, and the red line describes the flow path to the holding coil (HC).

Position B:
The sample port (2) is selected, and the fresh sample is drawn into the SI manifold, and is carried via the blue line to the HC.

Position C:
One of the other 8 ports is selected; these ports allow for sample, reagents, and other SIA manifold devices to be connected to the HC. The green line illustrates one example of this. This valve is for use in process analyzers where the sample port is coupled to a sampling system fast loop and provides fresh sample to a SIA analyzer. Other versions of the valve are available with additional flushed sample ports for multi-stream sampling.

 

© Global FIA, Inc, 2003
253-549-2223
800-581-6298 (TOLLFREE)
253-549-2283
info@GlobalFIA.com

Home | About Us | Buy | Tutorial | Library | What's New | Database | Contact Us | Search | Checkout